SECTION 15 n.

CALIBRATE BASE UNIT CONT.

3. Power on a DC current source capable of delivering up to

20.000mA and allow it to warm up to its specified accuracy as

defined by the manufacturer

4. Power on the HHC using either USB or battery power and

allow it warm up to its specified accuracy (5 minutes)

5. Set the current source to 0.000 mA

6. With the HHC in measure mode press the Ma/V key repeat-

edly until a mA reading appears on the display

7. Make note of the current source input current mA value and

the current mA value displayed on the HHC.

8. Verify that displayed current is within ±6uA (±0.006 mA)

9. Increase the input current setting from the current source at

desired intervals until reaching 20.000 mA.

10. Note input standard current and HHC displayed current and

calculate error.

11. It is recommended to at minimum verify display current at

0.000 and 20.000 mA input.

12. When test is complete, set current source to 0.00 mA and

disconnect the leads from the HHC.

SENSOR MODULE INPUT, 0 to +1.00000 V
WARNING Applying more than 1.25 volts to the calibration

module input could cause damage to the input of the HHC

NOTE

Sensor Module Calibration verification is performed

through Channel 1 module bay. Calibration values for

both module bays are verified and programmed via Chan-

nel 1. Functionality of Channel 2 bay is verified by install-

ing a pressure or temperature module.

NOTE

Any module inserted into the CH2 module bay during this

procedure could interfere with the resolution of the voltage

reading coming from the calibration module.

NOTE

If this process is performed following System Field Cali-

bration, damping is automatically set to 16 (maximum.).

To reduce the damping level to 1 (minimum), power cycle

the HHC prior to reading “As Left” data.

NOTE

The HHC will not recognize the calibration module as a

pressure or temperature module.

1.

Remove any pressure or temperature modules from both

system module bays.

2.

Insert the Calibration Module into the Channel 1 module bay.

3.

Leave Channel 2 bay empty or insert a system protection

module.

4.

Turn on the precision voltage standard and allow it to warm

up to its stated accuracy

5.

Set the voltage standard to a range to generate 0.00000v

6.

Connect the negative lead of the voltage source to the black

terminal on the Calibration Module

7.

Connect the positive lead of the voltage source to the red

terminal on the Calibration Module

8.

Power on the HHC using either USB or battery power and

allow it warm up to its specified accuracy (5 minutes)