4 delta sampling – Campbell Scientific PST3/PST8 Pump and Slug Test System User Manual

Page 23

Advertising
background image

SECTION 4. TECHNICAL DETAILS

4-2

TABLE 4.2-1. Keller-PSI Calibration Report

Customer: Campbell Scientific

Test Date:

07-06-92

Model No: 169-110-0051

Text Excitation

.5 mA

Serial No: 93907

Test Temperatures:

Room = 27ºC

Pressure Range: 0 to 15 PSIG

Cold = 0ºC

Excitation: .5 mA

Output: 0-60 mV

Test

BFSL

------- Run #1 -------

------- Run #2 -------

------- Run #3 -------

Pressure

Rm Temp

Rm Temp

Error

Rm Temp

Error

Cd Temp

Error

PSIG

Outputs

Outputs

%FSO

Outputs

%FSO

Outputs

%FSO

0.00

0.14

0.176

0.076

0.188

0.101

0.066

-0.006

3.00

9.55

9.519

-0.066

9.522

-0.060

9.449

-0.008

6.00

18.96

18.921

-0.083

18.916

-0.094

18.885

-0.006

9.00

28.37

28.358

-0.025

28.352

-0.038

28.350

-0.002

12.00

37.78

37.801

0.045

37.789

0.020

37.821

0.003

15.00

47.19

47.222

0.069

47.212

0.048

47.265

0.006

12.00

37.78

37.806

0.056

37.796

0.035

37.826

0.004

9.00

28.37

28.361

-0.019

28.356

-0.030

28.352

-0.001

6.00

18.96

18.926

-0.072

18.925

-0.074

18.888

-0.006

3.00

9.55

9.528

-0.047

9.528

-0.047

9.454

-0.008

0.00

0.14

0.189

0.103

0.189

0.103

0.073

-0.005

Maximum Static Error: 0.103 %FSO

Maximum Thermal Error @ Cold:

-.008 %FSO/C

Maximum Non-Repeatability: -0.26 %FSO

Electrical Termination: Black

+Input

Red

+Output

Green

-Output

White

-Input

With 3 wells or more on one PST, the first step
interval will be greater than 0.25 seconds (see
Section 4.4). The elapsed time will be offset by
the increased time required to execute the first
step interval. A change of the execution interval
to 0.5 seconds would result in the following:

Points per

Elapsed Time

Interval

Interval

0 - 40 sec.

0.5 sec.

80

40 - 240 sec.

10 sec.

20

4 - 20 min.

1 min.

16

20 - 200 min.

4 min.

45

200 - 2000 min.

20 min.

90

2000 min. plus

120 min.

2 times the value in *4 location 13 (PST3) or 28
(PST8)

If the interval is changed, it must be a multiple
of 0.125 (1/8) seconds. If the execution interval
were changed to 1 second, the following rate of
data collection would result:

Points per

Elapsed Time

Interval

Interval

0 - 80 sec.

1 sec.

80

80 - 480 sec.

20 sec.

20

8 - 40 min.

2 min.

16

40 - 400 min.

8 min.

45

400 - 4000 min.

40 min.

90

4000 min. plus

240 min.

NOTE: Never set the base interval greater
than 1 second.

4 times the value in *4 location 13 (PST3) or 28
(PST8)

4.4 DELTA SAMPLING

Time and water level are recorded only when
the water level changes by a user-specified
Delta threshold. When using Delta sampling,
each well in the test must have its own Delta
threshold value set in the *4 Table. The default

Advertising