Campbell Scientific TDR100 Time Domain Reflectometry Systems User Manual

Page 16

Advertising
background image

TDR100

FIGURE 4-1. PCTDR waveform for CS610 in water.

The algorithm will use the length of the waveform set by the Waveform
Length
. After finding the probe beginning, the algorithm searches over the
remaining waveform for the end of the probe. The length must be large
enough to display a short distance past the end of the probe under the wettest
expected conditions.

TABLE 4-1. Recommended Waveform Length values for

range of TDR probe rod lengths assuming soil porosity of

0.60.

Probe rod length (m)

Recommended Waveform Length

value (m)

0.10 to 0.20

3

0.21 to 0.30

4

0.31 to 0.40

5

0.41 to 0.6

6

0.61 to 0.75

7

0.76 to 1.00

9


Use the recommended values listed in Table 1 or use the following equation to
estimate the required window length, L

w

.