1: If Flag/Port (P91)
1: 21

Do if Flag 1 is Low

2: 0

Go to end of Program Table


;Measure the datalogger power supply voltage:

2: Batt Voltage (P10)
1: 1

Loc [ Bat_Volt ]


;Measure the internal temperature of the datalogger:

3: Internal Temperature (P17)
1: 2

Loc [ DL_Temp ]


;Set port 5 high to switch ON "Switched 12V" power from the datalogger to power the TDR100.

4: Do (P86)
1: 45

Set Port 5 High


;Measure the CS610 probe connected to Channel 3 of the level 1 SDMX50 and return La/L.

5: TDR100 Measurement (P119)
1: 0

SDM

Address

2: 0

La/L

3: 3001

MMMP Mux & Probe Selection

4: 4

Waveform

Averaging

5: 1

Vp

6: 250

Points

7: 9.5

Cable

Length

(meters)

8: 5

Window

Length

(meters)

9: .3

Probe

Length

(meters)

10: .085

Probe

Offset

(meters)

11: 3

Loc [ LaL_CH3 ]

12: 1.0

Mult

13: 0.0

Offset


;Square La/L to convert to dielectric constant:

6: Z=X*Y (P36)
1: 3

X Loc [ LaL_CH3 ]

2: 3

Y Loc [ LaL_CH3 ]

3: 4

Z Loc [ WC_CH3 ]


;Multiply the dielectric constant by 0.1 to prepare for the 3rd order polynomial
;with the appropriately scaled coefficients for Topp's calibration:

7: Z=X*F (P37)
1: 4

X Loc [ WC_CH3 ]

2: 0.1

F

3: 4

Z Loc [ WC_CH3 ]