Campbell Scientific TDR100 Time Domain Reflectometry Systems User Manual

Page 53

Advertising
background image

TDR100

;{CR10X}

;Program = T100EX#4

*Table 1 Program
01: 300

Execution

Interval

(seconds)


;Every 5 minutes (300 second execution interval)
;Measure the datalogger power supply voltage:

1: Batt Voltage (P10)
1: 1

Loc [ Bat_Volt ]


;Measure the internal temperature of the datalogger:

2: Internal Temperature (P17)
1: 2

Loc [ DL_Temp ]


;Once an hour on the hour (0 minutes into a 60 minute interval)
;Measure the TDR probes:

3: If time is (P92)
1: 0

Minutes (Seconds --) into a

2: 60

Interval (same units as above)

3: 30

Then

Do


;Set port 5 high to switch ON "Switched 12V" power from the datalogger to power the TDR100.

4: Do (P86)
1: 45

Set Port 5 High


;Measure the 6 CS610 probes connected to the Level 1 SDMX50
;(Channels 3 to 8) and convert to water content:

5: TDR100 Measurement (P119)
1: 00

SDM

Address

2: 0

La/L

3: 3006

MMMP Mux & Probe Selection

4: 4

Waveform

Averaging

5: 1

Vp

6: 250

Points

7: 5.75

Cable

Length

(meters)

8: 5

Window

Length

(meters)

9: .3

Probe

Length

(meters)

10: .085

Probe

Offset

(meters)

11: 11

Loc [ WC_3001 ]

12: 1138

Mult

13: -0.1758 Offset

;Switch the level 1 SDMX50 to Channel 1 and measure the 8 probes on the Level 2
;multiplexer that is connected to channel 1 of the level 1 mux (apparent cable length to
;probe 9.5 meters):

47

Advertising