6. Troubleshooting

If something unusual should occur, please check the following
items before you assume that there is a unit failure. For more
details, refer to the Troubleshooting section of the software
manual.

Scanning cannot be performed.

• Is the SCSI cable connected?

Connect the SCSI cable.

• Are you using the supplied SCSI cable?

Use the supplied SCSI cable

• Is the end of the connected SCSI cable terminated?

Connect the terminator.

• Is the SCSI ID number different from the SCSI ID of other

devices?
Change the ID number if it is duplicated.

• Are you using a recommended SCSI board ? (For Windows

version)
Use a recommended SCSI board.

• Are the numbers you have set for I/O, INT and DMA for

the SCSI board different from those for other boards? (For
Windows version)

• Does your system satisfy our recommended system

requirements?
Provide the recommended system.

• Have you installed software?

For information on installing and operating software, refer to
the software manual provided with the unit.

• Have you properly set the configuration, such as

CONFIG.SYS ?
If you have installed applications for Windows, for instance,
note that the configuration may change automatically. (For
Windows version)

Poor image quality.

• Have you set the film properly?

Set the film properly.

• Can you see dust on the film?

Remove dust on the film with a blower or photographic
brush.

• Was the unit subjected to vibration during the scanning

operation?
Place the unit where it will not be subjected to vibration or
physical shocks.

• Are you using utility software properly?

For details, please refer to the manual for your software.

The auto slide feeder display LED does not light.

• Is the LS-1000 power on?

Turn on the power.

• Are the LS-1000 and the auto slide feeder properly

connected?
Switch off the LS-1000 and re-connect.