Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

1

Installation

To convert from ESI, APCI or APPI to the multimode source

18 Connect the LC/MS sample tubing to the LC/MS diverter valve inlet filter.

See

Figure 19

on page 20.

Figure 19

LC/MS sample tubing connected to LC/MS inlet filter

19 If you are installing the multimode source for the first time, follow the steps

in

“Step 2. Install the HV control PCA and cables”

on page 9.

WA R N I N G

The LC/MS Liquid Chromatograph diverter valve is an integral part of the G1978B
safety system. The LC mobile phase flow must always be connected to the diverter
valve inlet filter. Never bypass the diverter valve and connect directly to the
nebulizer. If the diverter valve is used in a manner not specified by Agilent
Technologies, the protections provided by the diverter valve may be impaired.

LC/MS sample tubing