To do an autotune – Agilent Technologies G1978B User Manual

Page 32

Advertising
background image

32

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

2

Verification

To do an autotune

To do an autotune

Tuning the multimode source is done in MM-APCI mode only. Autotune is
done from the same menu as with all sources.

From the MSD Tune view, select Instrument > Autotune menu item.

The tune report will have a header with either the title MM-APCI Positive
Mode - Standard Scan
or MM-APCI Negative Mode - Standard Scan. You can
run the check tune after an autotune to validate that the instrument passes
check tune criteria. Do the autotune after the system has had at least 8 hours
to equilibrate vacuum and temperatures.

Figure 20

Autotune report