Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

2

Verification

To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes

To verify sensitivity for 6120 Quad - G1978B Multimode, Positive
SIM Modes

Preparation

1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric

flask. Use a clean graduated pipette.

2 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or

acetic acid.

3 Transfer 1 mL of the first dilution to a second 50 mL volumetric flask. Use a

clean graduated pipette.

4 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or

acetic acid.

This provides the final 2 pg/µL reserpine concentration required for
performance verification.

5 Transfer approximately 1 mL of the second dilution to a vial for use in the

LC autosampler.

N O T E

The G6110A instrument does not support the multimode source without the Selection
valve. Use the option kit G2735H to enable the G6110A to work with the multimode source.

Table 1

6110 and 6120 Quad Performance Verification Summary Table

G1978B
ES Positive SIM Mode

G1978B
APCI Positive SIM Mode

Sample

Reserpine, 5 ng/µL

Reserpine, 5 ng/µL

Concentration after dilution

2 pg/µL

2 pg/µL

Injection volume

5 µL

5 µL

Total sample amount injected

10 pg

10 pg

Sample order number

G2423A

G2423A

Solvent

75:25
methanol/water

with 0.1% formic or acetic acid

75:25
methanol/water

with 0.1% formic or acetic acid