C - Command Summary

84

CALCulate:LIMit:UPPer

This command sets various limits associated with rms voltage fluctuations testing for IEC 1000-3-3. as
described in the following table. All five parameters are type NRf. The order in which the five
parameters are entered must correspond to the order in the table.

vss delta

Sets the maximum peak-to-peak variation of relative voltage

1

that defines

“steady-state”. At *RST this value is set to 0.003

2

. Note that this number is

not specified by IEC 1000-3-3.

dmax limit

Sets the maximum relative voltage

1

change allowed before a dmax error is

flagged. At *RST this value is set to 0.04 (see note

2

).

dc limit

Sets the maximum relative steady-state voltage change allowed before a dc
error is flagged. At *RST this value is set to 0.03 (see note

2

).

dt limit

Sets the maximum time in seconds that the relative voltage

1

can exceed the

dt limit before a dt error is flagged. At *RST this value is set to 0.2 seconds.

dt limit

Sets the maximum relative voltage

1

that must be exceeded for dt limit

seconds before a dt error is flagged. At *RST this value is set to 0.03 (see
note

2

).

1

The expression “relative voltage” as used above is the measured rms voltage divided by the programmed

voltage.

2

This value is the ratio with respect to U

n

(the European nominal line voltage). For example, a value of

.03 represents 6.9 volts if U

n

= 230 volts. (Ratio * 100 = % of U

n

)

Command Syntax

CALCulate:LIMit:UPPer[:DATA] <NRf>,<NRf>,<NRf>,<NRf>,<NRf>

Parameters

see table

*RST Value

see table

Examples

CALC:LIM:UPP .003, .04, .03, .2, .03

Returned Parameters

<NR3>