E.

ELECTRICAL INSTALLATION

The electrical installation procedures provided in this manual are

common to all pH & ORP Monitor/controllers. See figure II.B.1 for

the hole dimensions of the enclosure’s cable access holes.

Unless otherwise instructed, refer to figure II.E.1. for the 720

Series II Monitor’s terminal block connector wiring designations.

NOTE: After removing an enclosure’s access hole plug, it is

suggested that the user mount a watertight restraint fixture prior

to installing a cable.

A device to disconnect the Model 720II from the

power supply is required. It is recommended that

this switch or circuit breaker be labeled as the

disconnection device for the Model 720II.

1. MAIN INPUT POWER INSTALLATION

WARNING: All Monitor/controllers are factory set for

115 VAC. Before starting, ensure the input power

“115/230” selection is correct for your requirements.

Failure to do so is beyond the responsibility of the

Myron L Company. See section II.E.2 below and

figure II.E.1.

NOTE: Some models may have either a 24 VAC or a

24 VDC input power requirement - check labels

carefully.

1. Verify that the main power source is turned “OFF” or

disconnected.

2. Using a standard slot screwdriver remove the

screws on the front panel.

3. Carefully wiggle the front panel to loosen the gasket and

pull gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

4. Turn the front panel around so that the back side is

facing you and set aside for now.

5. Carefully remove front panel, leaving the harness

connected. For OEM models skip to step #4.

6. Using the enclosure cutouts, install the proper wire and

watertight cable restraint (not provided) to comply with

local electrical codes. Skip for OEM.

7. Neatly connect wires to the Monitor/controller’s

connectors, as shown in figure II.E.1.

*CAUTION: The input power connectors require only a small

screwdriver or a pen to push on the release levers. The release

levers may be broken or damaged if not pushed straight toward

the circuit board. DO NOT push the release levers sideways.

2. 115/230 VAC CONVERSION

1. Before turning power on to the Monitor/controller ensure

the proper input voltage has been selected. Failure to do

so will blow the fuse. It could, under some conditions,

cause injury and damage the instrument voiding the

warranty.

2. Locate switch located next to the fuse holder.

3. Using a screwdriver, turn switch to required voltage.

3. CONNECTING THE SENSOR CABLE

For OEM models skip #1.

A 10 ft/3 meter sensor cable length is standard. 25 ft/7.5 meter and

100 ft/30 meter cable lengths are available as options. The sensor

cable may be extended up to 1000 feet/300 meters by ordering

cable only. It is recommended proper water tight and shielding

methods (i.e. junction box) be used when extending the cable.

Cable only: Part Number pHCRD-(length)

Example: pHCRD-300 (90m)

NOTE: 1000 ft/300 meter cable lengths have been tested with

no adverse affects on the readings. However, unusual conditions

may be experienced when extended long distances, i.e. power or

ground reversal. These conditions are beyond the control and

responsibility of the Myron L Company. A clean installation is

required. If you are experiencing errors, check the

Monitor/controller with a 10 ft/3 meter cable before contacting your

distributor or the Myron L Company.

1. Place the sensor cable and user supplied watertight

cable restraint into the enclosure’s appropriate access

hole.

2. Install the sensor cable wire to comply with local

electrical codes. Follow the color code as marked. See

figure II.E.1.

CAUTION: The circuit board connectors require only a small

screwdriver or a pen to push on the release levers. The release

levers may be broken or damaged if not pushed straight toward

the circuit board. DO NOT push the release levers sideways.

a. MODIFICATION FOR US PHARMACEUTICAL

25 (No Temperature Compensation)

This simple modification will allow your Monitor/controller to meet

the USP 25 requirements by disabling the normal temperature

compensation circuit thus giving “uncompensated” readings as

required.

12

!

WARNING !

!

CAUTION - READ FOLLOWING CAREFULLY

ELECTRICAL CONNECT DIAGRAM

CHASSIS GROUND for

OEM INSTALLATIONS ONLY

SENSOR

L N

0-10VDC

OUTPUT

NEU

GRN

RED

WHT

BLK

(+)

(-)

ALARM

CONTROL

RELAY

COM

NO

NC

MAIN

INPUT

POWER

GND-GRN
NEU-WHT
/ -DC
LINE-BLK
/ +DC

{

}

{

}

FUSE

115/ 230

SWITCH

Figure II.E.1