Myron L 729II User Manual

Page 25

Advertising
background image

The following instructions describe the steps for converting the

Monitor/controller. Refer to figure III.A.4 for the location of the

SPC(s) on your Monitor/controller. If unnecessary, skip to Set

Point Adjustment.

1. Ensure power is OFF.

2. Locate the SPC jumpers for the alarm to be configured,

see figure III.A.4.

3. Make a note of the current orientation of the SPC

jumpers.

4. Remove both jumpers. This is easily done by hand. Take

care not to crush the jumpers if using pliers.

5. Rotate the jumpers 1/4 turn (90°) and reinstall them on

their posts.

6. Continue with Set Point adjustment.

b. Set Point Adjustment

Set point #1 (LOW) must be adjusted BEFORE adjusting set

point #2 (HIGH).

1. Turn power ON.

2. While depressing the lower “SET POINT” switch, turn the

Set Point #1 control adjustment, figure III.A.4 until the

desired set point value is indicated on the display.

3. Repeat for Set Point #2 by depressing the upper “SET

POINT” switch and adjusting the SP2 on the SC/SCO

Module as shown in figure III.A.4.

4. Turn power OFF.

5. Continue or reinstall the front panel and tightly secure

retaining screws, see REASSEMBLY, page 22.

21

Figure III.A.4

SOLID STATE (24VDC 30mA) Output

PA™ PIEZO ELECTRIC ALARM OR

RA™ REMOTE ALARM OR

CUSTOMER CONNECTION

REMOVE TO

INSTALL

SET POINT #1

CONVERSION

FUSE*

115/

230

SET POINT #2 (high)

ADJUST

SC/SCO

OPTIONAL

SECOND

ALARM/

CONTROL

MODULE

SET POINT #2

CONVERSION

#2 RELAY

LIGHTS &

SWITCH

CONNECTOR

TRANSFORMER

721 726

722 727

723 728

724 729

MYRON L

COMPANY

RUBBER TAPE

(DO NOT

REMOVE)

RELAY #1

}

COM

GROUND

NC

NO

SPC

INC

PA

COM

RELAY #2

}

NO

NC

DIS

ORP ZERO

SOLID STATE (24VDC 30mA) OUTPUT

PA™ PIEZO ELECTRIC ALARM OR

RA™ REMOTE ALARM OR

CUSTOMER CONNECTION

SET POINT #2 HYSTERESIS

RIGHT INCREASING

LEFT DECEASING

3S

INC

SPC

SPAN

ZERO

HYS1

SP1

ZERO

DO NOT

ADJUST

0-10VDC

OUTPUT

SET POINT #1 (low)

ADJUST

SET POINT #1

HYSTERESIS

RIGHT INC

LEFT DEC

7 pH/ZERO mV

PUSH TO TEST

SPAN (GAIN)

PUSH TO TEST

UP

pH

CONFIGURATION

MODULE

}

{

POWER

AC LINE/ +DC
AC NEUTRAL/ -DC

BK WT RD GN NU R- R+

PWR

C GD NC NO CM

CHS

GND

PA

RA

CM

N

O

N

C

pH

}

SP

AN

DEC

RE

MO

VE

T

O

SE

CO

ND

R

EL

AY

IN

ST

AL

L

Advertising
This manual is related to the following products: