Dvm self - powered nc remote - powered – Myron L 729II User Manual

Page 29

Advertising
background image

requirements, i.e. self-powered or remote-powered, see

figure III.B.5.

3. Turn power ON

4. Press the SPAN (GAIN) Test Switch with the front panel

display at 0 pH (-1999 mV for ORP) the DVM should

indicate 4 milliamps.

5. If not, adjust the cal control marked “4mA” until the DVM

indicates 4mA, see figure III.B.3.

6. Press the ZERO Test Switch, with the display indicating

7 pH (0 mV for ORP) the DVM should indicate 12

milliamps.

7. If not, adjust the cal control marked “20mA” until the

DVM indicates 12mA. See figure III.B.3.

8. Calibration is complete.

9. Turn power OFF.

10. Reconnect sensor wires to sensor terminal block as

shown in figure III.B.4.

11. Carefully reinstall the front panel, bottom first, ensure no

wires have been pinched between enclosure and front

panel.

12. Reinstall the screws and tighten.

13. To operate, turn power ON.

25

Figure III.B.4

INC

DIS

ORP ZERO

CHS

GND

REMOVE TO

INSTALL

FUSE*

115/

230

TRANSFORMER

0-10VDC

MAIN CIRCUIT BOARD ASSEMBLY

721 726

722 727

723 728

724 729

MYRON L

COMPANY

3S

DEC

SPC

CONFIGURATION

MODULE

SPAN

ZERO

HYS1

SP1

SPAN

ZERO

DO NOT

ADJUST

7 pH/ZERO mV

PUSH TO TEST

SPAN (GAIN)

PUSH TO TEST

UP

}

ORP

4-20

CONNECTOR

CONNECT 4AO

HARNESS HERE

SENSOR

ORP

}

PWR

C GD NC NO CM

CHS

GND

RE

MO

VE

T

O

SE

CO

ND

R

EL

AY

IN

ST

AL

L

DEC

SP

AN

BK WT RD GN NU R- R+

PA

RA

(+)

(-)

+

-

(+)

SIGNAL OUT

POWER OUT

POWER IN

+

DVM

Self - Powered

NC

Remote - Powered

(+)

(-)

+

-

(+)

SIGNAL OUT

POWER OUT

POWER IN

+

DVM

Remote

Power

Supply

NC

Figure III.B.5

SO

P

O

P

I

SO

P

O

P

I

Advertising
This manual is related to the following products: