Absolute mode – In-Situ Tube 300R Telemetry System User Manual
Page 41
Advertising

800-446-7488
41
www.in-situ.com
When a Warning is triggered, the measurement is stored in the log memory even if the
Tube is not designated to save that reading. When an Alarm is triggered, the reading is
stored and an alarm message is sent (if the corresponding alarm has been enabled).
Alarm Mode is set in the Probe and Set-Point Configuration Menu, under the
Parameters Monitoring tab.
Absolute Mode
In Absolute Mode, alarms and warnings occur when a measurement value is less than
the minimum threshold value or greater than the maximum threshold value.
Minimum Level Value
Maximum Level
Value
Measurement Value
Result
1
15
16
ALARM
1
15
1
OK
1
15
15
OK
1
15
0
ALARM
-2
10
1
OK