Cleaning

Standard Cleaning

1.

Screw the instrument onto an empty bottle and empty it com-
pletely by dispensing (Fig. 1). If the instrument is equipped with
SafetyPrime

recirculation valve, it must be emptied in both the

"dispensing" and “recirculating” settings.

2.

Screw the instrument onto a bottle filled with a suitable cleaning
agent (e.g. deionized water) and rinse the instrument several
times by completely filling and emptying it.

3.

If the instrument is equipped with a SafetyPrime

recirculation

valve, after rinsing the instrument, it must also be rinsed in the
“recirculating” setting (Fig. 3).

Cylinder, valves, telescoping filling tube and discharge tube contain reagent. Point the
valves and tube openings away from your body. Wear protective clothing, eye protection
and appropriate hand protection.

Warning!

before the first use

immediately when the piston

is difficult to move

before changing the reagent

prior to long term storage

prior to dismantling the instrument

The instrument must be cleaned in the following situations to assure correct operation:

prior to changing the valves

regularly when using liquids which form

deposits (e.g., crystallizing liquids)

regularly when liquids accumulate in the

screw cap

The instrument should not be autoclaved!

Cleaning for trace analysis see page 42.