Troubleshooting, English – BrandTech Scientific Dispensette TA User Manual

Page 45

Advertising
background image

English

45

Liquid is above the
piston

Piston is leaking.

Perform a standard cleaning, replace the piston/
cylinder assembly (see page 39-41).

Piston difficult
to move

Formation of crystals, dirty

Stop dispensing immediately. Follow all cleaning
instructions (page 39-41).

Filling
not possible

Volume adjusted to
minimum setting

Set to required volume (see page 33).

Filling valve stuck

Clean the filling valve.If the valve is stuck use a
200 µl pipette tip to loosen it (see page 40). If
necessary, replace the filling valve and sealing
washer.

Dispensing
not possible

Discharge valve stuck

Clean discharge valve. If necessary replace dis-
charge tube (see page 40).

Air bubbles in
the instrument

Reagent with high vapor
pressure has been drawn in too
quickly

Draw in reagent slowly.

Seal not inserted, knurled
locking nuts not firmly
connected

Confirm that the seal is in place, and that the
knurled locking nuts at the discharge tube and the
SafetyPrime

recirculation valve are firmly seated

and finger tight.

The instrument has not
been primed

Prime the instrument (see page 32).

Filling tube is loose or
damaged

Push the filling tube on firmly. If necessary, cut

off approx. 1 cm of tube at the upper end and re-
connect it or replace filling tube.

Valves not firmly connected
or damaged

Follow cleaning procedure (see page 39, 42).
Tighten the valves using the mounting tool. If nec-
essary, replace the valves and sealing washers.

Recirculation tube not
connected

Connect recirculation tube (see page 31, Fig. 5).

Dispensed volume
is too low

Discharge tube is loose

Push the discharge tube on firmly and finger
thighten locking nuts.

Filling tube is loose
or damaged

Cleaning procedure (see page 39-40). Push the
filling tube on firmly. If necessary, cut off approx.
1 cm of the tube at the upper end and re-connect
it or replace filling tube (see page 40).

Filling valve is loose, con-
taminated, or damaged

Cleaning procedure (see page 39-40). Tighten
the valves using the mounting tool. If necessary,
replace filling valves and sealing washers.

Liquid leaking at the
safety ring

Dispensing cartridge is loose,
or the piston seal is damaged

Tighten the safety ring, and if necessary replace
the dispensing cartridge

Leaking liquid
between instrument
and bottle

Recirculation tube not con-
nected

Connect recirculation tube (see page 31, Fig. 5).

Volatile reagent dispensed
without sealing ring

Mount sealing ring.

Problem Possible cause Corrective action

Troubleshooting