38

First Steps

Is everything in the package?
Confirm that your package includes:

Bottle-top dispenser Dispensette

®

, discharge tube, telescoping filling tube, SafetyPrime

recirculation

valve and recirculation tube (optional), mounting tool, different bottle adapters, O-ring FKM (for autoclav-
ing), a performance certificate and this operating manual.

2.

Mounting the

SafetyPrime

recirculation valve
(optional)

3.

Mounting the

Discharge tube

Wear protective clothing, eye protection and gloves! Follow all safety
instructions and observe limitations of use and operating limitations
(page 32-34).

Nominal volume, ml

Adapters
bottle thread, PP

Filling tube
Length, mm

0,5

24-25, 28/ S 28, 32-33, 38 mm

125-240

1, 2, 5, 10

28/ S 28, 32-33, 38 mm

125-240

25, 50, 100

32-33, 38 mm

170-330

Warning:

Assembly

Push the SafetyPrime

recirculation

valve approx. 2 mm into the discharge
tube and firmly finger-tighten the locking
nut (Fig. 2). Check the SafetyPrime

recirculation valve for a tight fit.

Push the discharge tube (optional with
SafetyPrime

recirculation valve, Fig. 3')

approx. 2 mm into the valve block and
and firmly finger-tighten the locking nut
(Fig. 3). Check the discharge tube for a
tight fit.

Before mounting the SafetyPrime

recir-

culation valve or the discharge tue make
sure that the sealing washer is inserted.

1.

Check sealing

washer(s)

Recirculation valve and discharge tube
must correspond to the model. Pay
attention to the color code. On the next
day firmly tighten up the locking nut.

Note: