background image

!

20 °C

Ex

Type Digital • Easy
Calibration is manufac-
tured under U.S. Patent
5,957,330.

English

45

Error Limits

Error limits related to the nominal capacity (= maximum volume) indicated on
the instrument, obtained when instrument and distilled water are equilibrated
at ambient temperature (20 °C/68 °F). Testing takes place according DIN EN
ISO 8655-6 with a completely filled instrument and with uniform and smooth
dispensing.

Note:

* A = Accuracy, CV = Coefficient of Variation

Error limits Dispensette

®

The error limits are well within the limits of DIN EN ISO 8655-5. The maxi-
mum error limit for a single measurement can be calculated EL = A + 2 CV
(e.g. for volume 25 ml: 125 µl + 2 x 25 µl = 175 µl).

Nominal volume
ml

A*

±

%

µl

CV*

%

µl

0,5

1.0

5

0.2

1

1

0.5

5

0.1

1

2

0.5

10

0.1

2

5

0.5

25

0.1

5

10

0.5

50

0.1

10

25

0.5

125

0.1

25

50

0.5

250

0.1

50

100

0.5

500

0.1

100

Partial volume
The percentage values for A and CV are relative to the nominal volume (V

N

)

and must be converted for partial volumes (V

T

).

V

N

V

T

A

T

= · A

N

e.g.

volume

A*

±

%

µl

CV*

%

µl

V

N

25.0

0.5

125

0.1

25

V

T

= 50% N

12.5

1.0

125

0.2

25

V

T

= 10% N

2.5

5.0

125

1.0

25

* A = Accuracy, CV = Coefficient of Variation