Priming

Note:

Instruments without SafetyPrime

recirculation valve:

1.

Open the screw cap of the discharge tube (see instrument with

SafetyPrime

recirculation valve Fig. 1).

To avoid splashes, hold discharge tube orifice on the inner wall of
a suitable receiving vessel.

2.

For priming pull up the piston approx. 30 mm and push it down

rapidly until the lower stop.
Repeat this procedure approximately 5 times until the discharge
tube is bubble-free (Fig. 6).

Instruments with SafetyPrime

recirculation valve:

1.

Open the screw cap of the dispensing tube (Fig. 1). For safety,

hold the discharge tube orifice on the inner wall of a suitable
receiving vessel.

2.

Set valve to 'Recirculate' (Fig. 2).

3.

For priming gently pull up the piston approx. 30 mm and push it

down rapidly until the lower stop. Repeat this procedure 5 times
(Fig. 3).

4.

Turn valve to 'Dispense' (Fig. 4).

5.

To avoid splashes when priming hold the discharge tube on the

inner wall of a suitable receiving vessel and dispense liquid to
prime the discharge tube until it is bubble-free. Wipe away any
remaining drops from the discharge tube. (Fig. 5).

Before using the instrument for the first time, ensure it is rinsed
carefully and discard the first few samples dispensed. Avoid splashes.

Never press down the piston when the screw cap is screwed on!
Avoid splashing the reagent! The reagent can drip out from the dis-
charge tube and screw cap.