Commissioning ... / 7. maintenance, cleaning, Maintenance and cleaning – WIKA DIH10-Ex User Manual

Page 20

Advertising
background image

GB

20

WIKA operating instructions model DIH10

14089810.01 03/2014 GB/D/F/E

6.3.3 Display information overrange/underrange

Range:

4 ... 20 mA

Displayed range:

3.9 ... 20.1 mA

Usable range:

3.6 ... 21.5 mA

Advance warning of underrange: 3.6 ... < 3.9 mA

Advance warning of overrange:

> 20.1 ... 21.5 mA

Underrange displayed:

< 3.6 mA

Overrange displayed:

> 21.5 mA

With the advance warning, the display blinks (normal display

alternates with horizontal bars).

For values under 3.6 mA, a horizontal bar alternates with the display

“undr”. For values over 21.5 mA, a horizontal bar alternates with the

display “over”.

7. Maintenance and cleaning

7.1 Maintenance

The instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.

7.2 Cleaning

CAUTION!

■

Before cleaning, correctly disconnect the instrument

from the pressure supply, switch it off and disconnect it

from the mains.

■

Clean the instrument with a moist cloth.

■

Electrical connections must not come into contact with

moisture.

■

Wash or clean the dismounted instrument before

returning it, in order to protect persons and the

environment from exposure to residual media.

■

Residual media in dismounted instruments can result in

a risk to persons, the environment and equipment. Take

sufficient precautionary measures.