Next Advance SP1070-M User Manual

Page 33

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1000 series operator’s manual

Manual SP1000 series

26

Phase

Function

3

LP:ST

Phase

Function

4

LP:ST

Phase

Function

5

PS:90

Phase

Function

6

LP:03

Phase

Function

7

BEEP

Phase

Function

8

PS:30

Phase

Function

Rate

Volume

Direction

9

RATE

750 ml/hr

2.25 ml

Infuse

Phase

Function

Rate

Volume

Direction

10

RATE

750 ml/hr

0.25 ml

Withdraw

Phase

Function

11

LP:EN

9.4.3 Example 3: Ramping the Flow Rate

m

l/h

r

R

at

e

Time

0

150

200

250

Phases
2 - 4

Phases
5 - 8

Phases
9-11

Note: Graph is representative only.
Pumping rate increments and decrements in steps.

Continuously ramp up and down the pumping rate. Starting at 200 ml/hr, the pumping rate will increment
to 250 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Then the pumping rate will
decrement to 150 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed. Finally, the pumping rate
is incremented back to 200 ml/hr in 1.0 ml/hr steps after every 0.1 ml has been dispensed, then the process
is repeated.

Phase

Function

Rate

Volume

Direction

1

RATE

200 ml/hr

0.1 ml

Infuse

Phase

Function

2

LP:ST

Phase

Function

Rate

Volume

Direction

3

INCR

1.0

0.1 ml

Infuse

Phase

Function

4

LP:50

Phase

Function

5

LP:ST

Advertising