SP1000 series operator’s manual

Manual SP1000 series

40

<rate units> => UM

=

l/min

MM

= ml/min

UH

=

l/hr

MH

= ml/hr

Applicable only with “Rate Data Functions”.

When setting the pumping rate, if the current Phase’s function is not ‘RATE’, then <rate units> is
not applicable.

While pumping, the pumping rate can only be set if the current Phase function is ‘RATE’ and the
next Program Phase’s function to be executed is not ‘INCR’ or ‘DECR’. Also, while pumping,
<rate units> can not be set.

The new pumping rate will only be stored in non-volatile memory if the Pumping Program is not
operating.

When the pumping rate is queried while pumping, the response will be the current pumping rate
and units. Otherwise, the response will be the rate setting and units, if applicable. With the
‘INCR’ and ‘DECR’ functions, these two responses are not the same.

RAT C <float> [ <rate units> ]

Allows the Pumping Program to continue after the pumping rate is changed. The RAT C
command will not clear a Paused Pumping state after the rate is changed when the Pumping
Program is paused. Normally, the Pumping Program would re-start from Phase 1 after a rate
changed while paused.

RAT I <float> [ <rate units> ]

Changes the pumping rate only if the pump is infusing. If the pumping direction is set to
withdraw, the rate change will be ignored.

VOLUME TO BE DISPENSED AND SET VOLUME UNITS

VOL [ <float> | <volume units>]

Set/query volume to be dispensed: <float>
Applicable only with “Rate Data Functions”. Can only be set when the Pumping Program is not
operating. The volume units are set according to the current syringe diameter setting. Do not send
the volume units when setting the volume.
Example: VOL 12.45

Sets the current phase “Volume to be Dispensed” to 12.45.

Set volume units: <volume units>
Overrides the default volume units set when the syringe diameter is set.
Example: VOL UL

Sets all volume units to

l.

Query response:

<float> <volume units>

PUMPING DIRECTION

DIR [ INF | WDR | REV ]

Set/query pumping direction

INF

= Infuse

WDR

= Withdraw

REV

= Reverse pumping direction

Applicable with all Program Phase functions. Cannot be set when the Pumping Program is
operating and the “Volume to be Dispensed” is non-zero.

The pumping direction cannot be changed if an alarm condition exists.

Query response: { INF | WDR }