Appendix, Ttl i/o control from the pumping program, Ttl i/o control from rs-232 – Next Advance SP1070-M User Manual

Page 55: Rs-232 command summary

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP1000 series operator’s manual

Manual SP1000 series

48

11.2

TTL I/O Control from the Pumping Program

Various Pumping Program functions can define how the pump reacts to levels on the TTL I/O connector or
set output levels. These are summarized in the following table:

Pumping
Program
Function

TTL I/O Control Pin

Pin #

Action

OUT.n

Program Output

5

Set logic level output to ‘n’

EV:nn

Event Trigger

4

Falling edge triggers a jump to Phase ‘nn’

ES:nn

Event Square wave
Trigger

4

Rising or falling edge triggers a jump to Phase 'nn'

IF:
Programmab
le Syringe
Pump nn

Program Input

6

Low level causes a jump to Phase ‘nn’

PS:00

Operational Trigger

2

Trigger activation resumes Program operation

11.3

TTL I/O Control from RS-232

The logic levels of pins 2, 3, 4, and 6 can be queried from an attached computer using the RS-232 ‘IN’
command.

The output logic level of pin 5 can be set with the RS-232 ‘OUT’ command.

12.

Appendix

12.1

RS-232 Command Summary

<command> =>
DIA [ < float > ]

Syringe inside diameter

PHN [ < phase data > ]

Program Phase number

FUN [ < phase function > ]

Program Phase function

< phase function > =>

RAT

Pumping rate. ‘RATE’

FIL

Fill syringe to volume dispensed. ‘FILL’

INC

Increment rate. ‘INCR’

DEC

Decrement rate. ‘DECR’

STP

Stop pump. ‘STOP’

JMP <phase data>

Jump to Program Phase. ‘JP:nn’

PRI

Sub-Program Selection Input. 'Pr:In'

PRL <number data>

Sub-Program Start Label definition. 'Pr:nn'

LOP <count data>

Loop to previous loop start ‘nn’ times. ‘LP:nn’

LPS

Loop starting Phase. ‘LP:ST’

LPE

Loop end Phase. ‘LP:EN’

PAS <number data>

Pauses pumping for ‘nn’ seconds. ‘PS:nn’

PAS [n.n]

Pauses pumping for 'n.n' seconds. 'PS:n.n'

IF <phase data>

If Program input low, jump to Program Phase. ‘IF:nn’

EVN <phase data>

Set event trigger. ‘Et:nn’

EVS <phase data>

Set event square wave trigger. 'ES:nn'

EVR

Event trigger reset. ‘Et:RS’

TRG <n

n

>

Override default operational trigger configuration ‘tr:aa’

BEP

Sound short beep. ‘BEEP’

OUT { 0 | 1 }

Set programmable output pin. ‘OUT.n’

RAT [C | I ] [ <float> [ UM | MM | UH | MH ] ] Pumping rate
VOL [ <float> | <volume units>]

Volume to be Dispensed, or set Volume units

DIR [ INF | WDR | REV ]

Pumping direction

RUN [ <phase data> ]

Starts the Pumping Program

[E [<phase data>] ]

Pumping Program event trigger

STP

Stop/pauses the Pumping Program

Advertising