Changing Volume Units
The volume units used for the accumulated volume can be changed to either ‘ml’ or ‘

l’. NOTE: Volume

units can only be changed immediately after setting the syringe diameter and before the pump is started.

After the syringe diameter is entered, display the “Volume Dispensed” by pressing the “Volume” key.
The current volume units and the “Dispensed” LED will be lit.

Pressing any up arrow key will change the display to

and the current volume units will blink.

Then, press any up arrow key to switch the volume units between ‘ml’ and ‘

l’. Press any non-arrow key

or wait 2 seconds to enter the new volume units. The display will blink when entered. The selected
volume units will remain in affect and override the default volume units. Changing the diameter will no
longer change the volume units. Performming a system reset will cancel the override and allow the
volume units to change to the default volume units when setting the syringe diameter.

6.2

Operating the Pump

When the “Start/Stop” key is pressed, the Pump will begin pumping, and the ‘Pumping’ LED will be lit.
The pump will pump continuously until stopped.

Either the pumping rate, “Volume Dispensed”, or syringe diameter can be displayed while pumping.
Press the relevant key to change the display. While displaying the pumping rate, press the up arrow keys
to change the pumping rate.

6.3

Purging

To purge the syringe, with the Pump stopped, press and hold the 'Start/Stop' key. The Pump will start
pumping, then, after one second, purge will begin. The pump will pump at its top speed. Purging will
continue until the 'Start/Stop' key is released, and then the pump will stop. While purging the display will

show:

.

Press and hold the direction button, to perfrom a reverse purge. The display will show:

6.4

Changing the Pumping Rate While Pumping

The pumping rate can be changed while the pump is operating. To change the pumping rate, display the
current pumping rate by pressing the ‘Rate’ key, if needed. With the pumping rate displayed, press the up
arrow keys to change the rate. The rate units will blink while the rate is being changed. Rate units can not
be changed while pumping.

The new rate is stored after a 2 second delay or by pressing a non-arrow key. If the new rate is within the
operating range of the pump, the display will blink and the new rate will be stored in memory and the
pump will begin pumping at the new rate. If the new rate is out of the operating range of the pump, the

display will show

. Pressing any key clears the message and restores the previous rate.

6.5

Volume Dispensed

The accumulated Volume Dispensed can be displayed at any time by pressing the Volume key. The
display will show the total accumulated volume pumped with the ‘ml’ or ‘

l’ LED lit and the ‘Dispensed’

LED lit. Volume is computed based upon the syringe inside diameter setting.

The “Volume Dispensed” accumulation is reset to 0 when:

A) With the pump stopped, pressing and holding any up arrow key while displaying the volume.
B) The syringe diameter is changed.
C) The accumulated Volume Dispensed rolls over from 9999 to 0.