Syringe diameters and rate limits – Next Advance SP300 User Manual

Page 14

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP300 Programmable Syringe Pump

Publication #1200-01

11

03/23/10

9.5

Syringe Diameters and Rate Limits

Next Advance

www.nextadvance.com

Model: SP300 syringe pump

Maximum and minimum flow rates

Syringe

Manufacturer

Syringe

(cc)

Inside

Diameter
(mm)

Maximum

Rate
(mL/hr)

Minimum

Rate
(µl/hr)

Maximum

Rate
(mL/min)

B-D

1

4.699

39.27

0.73

0.6545

3

8.585

131

2.434

2.184

5

11.99

255.6

4.748

4.261

10

14.43

370.3

6.876

6.172

20

19.05

645.4

11.99

10.75

30

21.59

829

15.4

13.81

60

26.59

1257

23.35

20.95

HSW

1

4.69

39.11

0.727

0.6519

Norm-Ject

3

9.65

165.6

3.076

2.76

5

12.45

275.6

5.119

4.594

10

15.9

449.6

8.349

7.493

20

20.05

714.9

13.28

11.91

30

22.9

932.6

17.32

15.54

50

29.2

1516

28.16

25.27

Monoject

1

5.74

58.59

1.088

0.9766

3

8.941

142.1

2.64

2.369

6

12.7

286.8

5.326

4.78

12

15.72

439.4

8.161

7.324

20

20.12

719.9

13.37

11.99

35

23.52

983.8

18.27

16.39

60

26.64

1262

23.44

21.03

140

38.00

47.69

42.80

Terumo

1

4.7

39.28

0.73

0.6547

3

8.95

142.4

2.646

2.374

5

13

300.5

5.581

5.009

10

15.8

443.9

8.244

7.399

20

20.15

722.1

13.41

12.03

30

23.1

949

17.63

15.81

60

29.2

1516

28.16

25.27

Air-Tite

10

15.9

449.6

8.349

7.493

20

20.25

729.2

13.55

12.15

30

22.5

900.3

16.72

15

50

29

1495

27.78

24.92

Syringe

(µL)

Inside

Diameter
(mm)

Maximum

Rate
(µl/hr)

Minimum

Rate
(µl/hr)

Hamilton

0.5

0.103

18.86

0.001

Microliter

1

0.146

37.91

0.001

2

0.206

75.47

0.002

5

0.326

189

0.004

SGE

0.5

0.1

17.78

0.001

1

0.15

40.01

0.001

Advertising