9.2

Troubleshooting and Maintenance

Maintenance: Periodically, a small amount of all-purpose oil should be applied to the guide rods.

The mechanism should be kept clean to prevent impeded operation.

No other special maintenance or calibrations are needed

Pusher block makes a snap or click sound when the pump is started: This is a normal condition.
When the pusher block is manually moved, the drive-nut may not have been fully engaged on the drive
screw. The sound heard is the drive-nut engaging on the drive screw.

9.3

Specifications

Mechanical & Electrical

Syringe sizes:

Up to 140 cc

Number of syringes:

1

Motor type:

Step motor

Motor steps per revolution:

400

Microstepping:

1/8 to 1/2 depending on motor speed

Advance per step:

0.2126 uM to 0.8504 uM depending on motor speed

Motor to drive screw ratio:

15/28

Drive screw pitch:

20 revolutions/”

DC connector:

2.1 mm, center positive

Voltage at DC connector:

12V DC at full load

Amperage:

750 mA at full load

Power supply type:

Unregulated linear external wall adapter, country and power
source specific

Power supply output rating:

12V DC @ 800 mA

Dimensions:

8 3/4” x 5 3/4” x 4 1/2” High
(22.86 cm x 14.605 cm x 11.43 cm)

Weight:

3.6 lbs. (1.63 kg)

Operational

Maximum speed:

3.7742 cm/min

Minimum speed:

0.004205 cm/hr

Maximum pumping rate:

1257 ml/hr with a B-D 60 cc syringe

Minimum pumping rate:

0.73

l/hr with a B-D 1 cc syringe

Maximum force:

35 lbs. at minimum speed, 18 lbs. at maximum speed

Syringe inside diameter range:

0.100 to 50.00 mm

9.4

Custom Applications

For specialized and OEM applications, contact your dealer
or

Next Advance

Inc. Custom modifications can be made to

the mechanics or the firmware.