- 42 -

• Place the required weight on the weighing platform and press

key.

The adjustment process starts automatically, which is signaled by the
displaying the following message:

• The completion of the adjustment procedure is signaled by the

following message:

• Take off the adjustment weight from the weighing platform, after which

message <donE> is displayed for approximately 1s and the scale
returns to displaying name of submenu: adjustment:

• Adjustment process can be aborted at optional moment by pressing

key, which is signaled by displaying the following message:

Caution:

1. Remember to carry out scale’s adjustment process when there is no

load on the scale’s weighing platform!

2. If the adjustment process (span adjustment) lasts longer than 15

seconds, error message <Err8> is displayed and short beep sound is

emitted. Press

key and repeat adjustment procedure under more

stable ambient conditions!