Caution:
If the start mass adjustment lasts longer than 15 seconds, error message
<Err8> is displayed and short beep sound is emitted. Press
key and
repeat adjustment procedure under more stable ambient conditions!
Return to weighing with procedure of saving changes:
See point - 11.2.2. – return to weighing.
18. COOPERATION WITH A PRINTER
Each time the
key is pressed causes sending a signal containing
current mass value and a measuring unit.
Depending on setting of STAB parameter, the printout is released with
stable or unstable (instantaneous) value. Depending on setting of REPL
parameter, printout is automatic or manual.
Medical scales enable cooperating with one of the following printers
KAFKA series:
a) KAFKA
Printout of weighing result and measuring unit only
b) KAFKA 1/Z
This printer features an internal real time clock. Both date and time
are printed on plugging the printer to mains and connecting it to the
scale.
c) KAFKA SQ S
This printer features an internal real time clock and possibility of
running statistics from measurements. Statistic contents: quantity of
samples, total mass of weighed samples, average value, standard
deviation, variation factor, min value, max value, difference max –
min.