2.8.1.8

Configuration Window

The Configuration window allows the user to set up the data collection and analysis parameters. See
Figure 26.

Figure 26. Data Collection Configuration Window

2.8.1.8.1 Number of Samples
The Number of Samples field allows the user to select the number of samples to collect, between 16 and
524288.

2.8.1.8.2 Average
When performing FFT analyses, the Average field determines the number of FFTs to average. FFTs will
be collected and averaged when the Collect button is pressed.

2.8.1.8.3 FFT Window
The FFT Window box allows the user to select the type of windowing algorithm for FFT processing.
Windowing algorithms include the Blackman, Blackman-Harris, Hanning, 5-term Hodie, and 7-term Hodie.
The 5-term Hodie and 7-term Hodie are windowing algorithms developed at Crystal Semiconductor, now
Cirrus Logic.

2.8.1.8.4 Histogram Bin Width
This field determines the "bin width" when plotting histograms of the collected data. Each vertical bar in
the histogram plot will contain the number of output codes entered in this field. Increasing this number
may allow the user to view histograms with larger input ranges.

2.8.1.8.5 Seconds to Delay
This field specifies the amount of time the system waits to begin data collection after the CS5480 starts
continuous conversions.

2.8.1.8.6 Data to Collect
The Data to Collect check boxes allow the user to select the data types that will be collected and returned
to the PC for processing.

2.8.1.8.7 Accept Button
When the Accept button is pressed, the current settings will be saved and the user will return to the Data
Collection
window.