Functional Verification: Gate In Test (cont’d)
Test Procedure
(cont’d)
4.
Set up the AFG to output a 1 MHz sinewave with an external gate
source:
TRIG:GATE:SOUR EXT;
External gate source
:TRIG:GATE:STAT ON;
Enable gate
:FREQ 1E6;
Set freq to 1 MHz
:VOLT 4VPP
Set AFG amplitude
INIT:IMM
Initiate
5.
Send the following command to the Command Module to set the
level at the "Trig Out" port to 5 V. Verify that the scope shows a 1
MHz sinewave.
OUTP:EXT:LEV 0
6.
Send the following command to the Command Module to set the
level at the "Trig Out" port to 0 V. Verify that the scope shows a DC
signal.
OUTP:EXT:LEV 1