Test 2-10: Spurious/Non-Harmonic Distortion
Description
The purpose of this test is to verify that the AFG meets its specifications for
non-harmonic and spurious distortion.
Equipment Setup
•
Connect equipment as shown in Figure 2-9
•
Set Spectrum Analyzer to :
Ref Level = -5 dBm
Resolution BW = 3 kHz
Video BW = 3 kHz
NOTE
These are recommended settings only. Adjust your Spectrum Analyzer as
necessary.
Test Procedure
1.
Reset the AFG:
*RST;*CLS
Reset AFG and clear
status registers
2.
Set the AFG to output a -5 dBm, 10 MHz sinewave with the 10 MHz
filter enabled:
FREQ 1.0E7;
Set AFG frequency
:VOLT -5DBM
Set AFG amplitude
OUTP:FILT:FREQ 10 MHZ
Set filter to 10 MHz
OUTP:FILT ON
Enable filter
INIT:IMM
Initiate waveform
Agilent E1445A Service Manual