DC Adjustment Procedure (cont’d)

Test Procedure
(cont’d)

5.

Trigger the DMM and note the reading.

6.

Send the reading to the AFG:

CAL:DC:POINT? <reading>

where <reading> is the DMM reading from step 5. The AFG will
return, in order, the number of the current cal point and an error code.
Any non-zero error code indicates a failure.

Table 3-1. DMM Range Setting for Cal Points 41 and 43

Absolute Value

of DC Offset

DMM Range

Less than 10 mV

100

Ω

Between 10 mV and 100 mV

1000

Ω

Greater than 100 mV

Offset too high - abort cal

Table 3-2. DC Calibration Points

Cal

Point

DMM Settings

(changes only)

1
29
31
33
41
43

DCV, 10 V range, 10 NPLC
1 V range, 20 NPLC
10 V*
100 mV range
4-wire ohms, offset comp on*
4-wire ohms, offset comp on*

*See Step 4 of the "DC Adjustment Procedure" for
special instructions.

Agilent E1445A Service Manual