Test 2-9: Total Harmonic Distortion

Description

The purpose of this test is to verify that the AFG meets its specifications for
sine wave total harmonic distortion (THD).

Equipment Setup

•

Connect equipment as shown in Figure 2-10

•

Set Spectrum Analyzer to:

Ref Level = 25 dBm
Freq Span = 1 kHz
Resolution BW = 30 Hz
Video BW = 30 Hz

NOTE

These are recommended settings only. Adjust your Spectrum Analyzer as
necessary.

Figure 2-10. Equipment Setup for Test 2-9 and Test 2-10