Test the ECL threshold

1

Select the field to the right of Pod 1, then select ECL in the pop-up menu.

2

On the function generator front panel, enter

−1.160 V ±1 mV DC offset. Use the

multimeter to verify the voltage.

The activity indicators for pod 1 should show all data channels and the J-clock channel at a
logic high.

3

Using the Modify down arrow on the function generator, decrease offset voltage in
1-mV increments until all activity indicators for pod 1 show the channels are at a
logic low. Record the function generator voltage in the performance test record.

4

Using the Modify up arrow on the function generator, increase offset voltage in
1-mV increments until all activity indicators for pod 1 show the channels are at a
logic high. Record the function generator voltage in the performance test record.

Testing Performance
To test the threshold accuracy