Test the + user threshold – Agilent Technologies 01664-97005 User Manual

Page 36

Advertising
background image

Test the + User threshold

1

Move the cursor to the field to the right of Pod 1. Type +6.00, then use the left and
right cursor control keys to highlight V. Press the Select key.

2

On the function generator front panel, enter +6.282 V

±1 mV DC offset. Use the

multimeter to verify the voltage.

The activity indicators for pod 1 should show all data channels and the J-clock channel at a
logic high.

3

Using the Modify down arrow on the function generator, decrease offset voltage in
1-mV increments until all activity indicators for pod 1 show the channels at a logic
low. Record the function generator voltage in the performance test record.

4

Using the Modify up arrow on the function generator, increase offset voltage in
1-mV increments until all activity indicators for pod 1 show the channels at a logic
high. Record the function generator voltage in the performance test record.

Testing Performance
To test the threshold accuracy

3–14

Advertising