Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup time of the setup/hold combination selected, +0.0 ps or

−100

ps.

a

In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set
Marker 1 at

−1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at

−1.3000 V.

b

In the oscilloscope Delta T menu, select Start On Pos Edge 1. Select Stop on Neg
Edge 1.

c

Adjust the pulse generator channel 1 Delay, then select Precision Edge Find in the
oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to
the setup time of the setup/hold combination selected, +0.0 ps or

−100 ps.

10

Select the clocks to be tested.

a Select the clock field to be tested, then select J

↓ + K↓ as the clock edges.

b

Select Done to exit the Master Clock menu.

Testing Performance

To test the multiple-clock, multiple-edge, state acquisition

3–43