Trouble-shooting

Error message

Situation

Solution

No lens filter is attached.

Attach new, clean lens filter.

The thermometer cannot identify
a secure position of the probe.
An accurate measurement was
not possible.

POS = position error

IRT 4020: Clear the display by
pushing the «start» button once.
IRT 4520: Clear the display by
pushing the «I/O» button once.
Take care that the positioning of
the probe is correct and remains
stable.

Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).

Allow the thermometer to remain
in a room for 30 minutes where
the temperature is between
10 and 40 °C or 50 and 104 °F.

Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2–108 °F).

HI = too high

LO = too low

Make sure a new, clean lens filter
is attached and thermometer is
properly inserted. Then, take a
new temperature.

System error – self-check display
flashes continuously and will not
be followed by the ready beep
and the ready symbol.

If error persists,

If error still persists,

Wait 1 minute until the thermo-
meter turns off automatically, then
turn on again.

... reset the thermometer by re-
moving the batteries and putting
them back in.

… have the thermometer checked
at a Braun Customer Service
Center.