PROGRAMMING THE BREWER (cont.)

Procedures to calibrate the bypass flow rate:

1. Place a container, accurately graduated in ounces,

and with a minimum capacity of 60 ounces, below

the funnel on the brewer.

2. Press and hold the upper right hidden switch until

the display reads UNITS.Release the switch, then

press and release switch until the display reads

CALIBRATE FLOW.

3. Press and release YES to advance to the SPRAY-

HEAD CAL? function screen. (Pressing NO in the

CALIBRATE FLOW screen will advance to the next

function screen, BREW COUNTERS).

4. Press and release NO to advance to the BYPASS

CALIBRATION screen.

ENTER OZ XX.X

(-) DONE (+)

NEW BYPAS FLOW

XX.X OZ/M

EXIT TO THE

NEXT FUNCTION

SCREEN

COUNTS DOWN

TO ZERO

BREW PRESSED

AFTER 1 MINUTE

CALIBRATE FLOW?

NO YES

CONTAINER RDY?

NO YES

CALIBRATE BYPASS

PRESS BREW START

CALIBRATE BYPASS

60 SEC TO FINISH

BYPASS CAL?

NO YES

SPRAYHEAD CAL?

NO YES

P2314

NEW SPRAY FLOW

XX.X OZ/M

ENTER OZ XX.X

(-) DONE (+)

CALIBRATE FLOW?

NO YES

SPRAYHEAD CAL?

NO YES

CONTAINER RDY?

QUIT YES

CALIBRATE SPRAY

PRESSBREWSTART

CALIBRATE SPRAY

60 SEC TO FINISH

COUNTS DOWN

TO ZERO

AFTER 1 MINUTE

BREW PRESSED

EXIT TO THE

NEXT FUNCTION

SCREEN

P2307

CALIBRATE FLOW (cont.)

5. Press and release the YES pad. The display should

read CONTAINER READY? If the container is

under the funnel, press YES. The display should

read CALIBRATE BYPASS .. PRESS BREW TO

START.

6. Press and release the BREW switch. The display

should read CALIBRATE BYPASS .. 60 SEC TO

FINISH. The 60 second timer on the display will

count down to zero. When the counter reaches

zero, the display will change to ENTER OZ., along

with a number measured in ounces per minute.