Environmental

Shelf Life: 2 years at 25° C (77° F)

Operating Temperature: -5 to 35° C (23 to 95° F) Storage Temperature: -20 to 60° C (-4 to 140° F)

Relative Humidity (operating & storage): 15 to 90% (non-condensing)

Physical

Dimensions: 5.5 x 5.5 x 1 inches (14 x 14 x 2.5 cm)

Weight: < 14 ounces (0.4 kg)

Standards Compliance

This device is designed to conform to the following standards:

IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment

IEC 60601-1-2 Electromagnetic Compatibility

IEC 60825-1:1993+A1:1997+A2:2001 (Class 1 LED Product)

IEC 60601-1 Classification

Type of Protection Against Electric Shock: Internally powered equipment

Degree of Protection Against Harmful Ingress of Water: IPX0 (Ordinary protection against ingress of liquids)

Mode of Operation: Continuous Operation

Not suitable for use in the presence of a flammable anaesthetic mixture with air, oxygen, or nitrous oxide.

Electrical

goLITE BLU DC Power Consumption: 9.0 V DC, 1.1 A

AC/DC Adapter Power Consumption: 100 – 240 VAC, 50/60 Hz, 0.4 A

AC/DC Adapter: (model number: MW170KB0900B02) 9.0 V DC, 1.5A
As of the date of publication, this is the complete list of approved AC/DC adapters for use with the goLITE BLU.
Contact customer service for an updated list of approved adapters.

Internal Battery: (part number: 1053538) 2-cell rechargeable lithium-ion pack, 7.4 V, 850 mAh

Complies with IEC 62133 and UL 2054

Disposal

Dispose of the device in accordance with local regulations.