Emc inf or ma tion – Philips HF3329-60 User Manual

Page 24

Advertising
background image

23

G

uiDance

anD

m

anufacturer

S

D

eclaration

- e

lectromaGnetic

i

mmunity

: The goLITE BLU is intended for use in the electromagnetic

environment specified below. The user of the goLITE BLU should make sure it is used in such an environment.

i

mmuniTy

T

EsT

iEC 60601 T

EsT

l

EvEl

C

omplianCE

l

EvEl

E

lECTromagnETiC

E

nvironmEnT

- g

uidanCE

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications equipment should be

used no closer to any part of the goLITE BLU, including cables,

than the recommended separation distance calculated from the

equation applicable to the frequency of the transmitter.

Recommended separation distance:

d = 1.2

150 kHz to 80 MHz

d = 1.2

80 MHz to 800 MHz

d = 2.3

800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter

in watts (W) according to the transmitter manufacturer and d is

the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined

by an electromagnetic site survey a, should be less than the

compliance level in each frequency range b.

Interference may occur in the vicinity of equipment marked with

the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and

people.

a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and

FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF

transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the goLITE BLU is used

exceeds the applicable RF compliance level above, the goLITE BLU should be observed to verify normal operation. If abnormal performance is observed,

additional measures may be necessary, such as re-orienting or relocating the goLITE BLU.

b: Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.

EMC