Emc inf or ma tion – Philips HF3329-60 User Manual

Page 23

Advertising
background image

22

G

uiDance

anD

m

anufacturer

S

D

eclaration

- e

lectromaGnetic

i

mmunity

: The goLITE BLU is intended for use in the electromagnetic

environment specified below. The user of the goLITE BLU should make sure it is used in such an environment.

i

mmuniTy

T

EsT

iEC 60601 T

EsT

l

EvEl

C

omplianCE

l

EvEl

E

lECTromagnETiC

E

nvironmEnT

-

g

uidanCE

Electrostatic Discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete

or ceramic tile. If floors are covered

with synthetic material, the relative

humidity should be at least 30%.

Electrical Fast Transient/Burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input-output lines

±2 kV for supply mains

±1 kV for input/output lines

Mains power quality should be

that of a typical home or hospital

environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV for common mode

Mains power quality should be

that of a typical home or hospital

environment.

Voltage dips, short

interruptions and voltage

variations on power supply

input lines

IEC 61000-4-11

<5% U

T

(>95% dip in U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

) for 5 cycles

70% U

T

(30% dip in U

T

) for 25 cycles

<5% U

T

(>95% dip in U

T

) for 5 sec

<5% U

T

(>95% dip in U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

) for 5 cycles

70% U

T

(30% dip in U

T

) for 25 cycles

<5% U

T

(>95% dip in U

T

) for 5 sec

Mains power quality should be

that of a typical home or hospital

environment. If the user of the goLITE

BLU requires continued operation

during power mains interruptions, it

is recommended that the goLITE BLU

be powered from an uninterruptible

power supply or a battery.

Power frequency (50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields

should be at levels characteristic of a

typical location in a typical hospital

or home environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

EMC

Inf

or

ma

tion