BNC 7000 Series User Manual User Manual

Page 24

Advertising
background image

24

A) Enable internal frequency counter and power meter (see Figure 6): If the auto checkbox is

enabled, the counter and power meter are running and constantly updating the detected DUT

frequency and power. Disabling allows manual entry of the DUT frequency. Internal counters

and power detector is shut down. Per default, the auto-detection is on and once a valid signal

is detected, the Measure button turns green and is enabled.

B) Set frequency offset range using slider.

C) Select the number of correlation: The correlation number can be set to predefined

values using drop-down menu or can be set to any user value by typing in the number.

D) Use the “More” Menu to select between internal or external references sources.

Per default, the internal sources are used and no additional configuration is needed.

See the “Using external references” section for more information how to use external

references.

Figure 7 Setup for phase noise measurement

E) Configure Continuous & Save Trace Options: With “Continuous” is enabled, the

measurement is restarted automatically after completing the selected number of correlations.

With “Save Trace” enabled, the trace of a completed measurement is automatically imported

and stored in the Trace List.

F) Click Measure Button: Once the Measure Button becomes green, a new measurement can

be started. Click on the button to start the measurement and click again to stop the

measurement before completion.