BNC 7000 Series User Manual User Manual

Page 42

Advertising
background image

42

1. Determine the phase detector constant. Start this step by pressing the “Measure” button in step1.

Mechanically or digitally tuned phase shifters must be manually adjusted to provide at least 180

degrees phase shift to derive the phase detector constant. If already known for a given setup, the

phase detector constant can be manually entered. In case “Analog Tune” is selected and voltage

controlled phase shifters are used, this step is performed automatically by the instrument and can

be omitted.

2. Once the phase detector constants are known, the phases must be adjusted to obtain phase

quadrature. Start this step by pressing the “Measure” button in step 2. Now, the phase difference

between signal pairs at RF IN and REF IN (1 and 2) must be adjusted using the external phase

shifters to achieve approximate phase quadrature (90 degree phase difference) as shown in

Figure 17. The GUI shows a tolerance window of ±3 degrees. In case “Analog Tune” is selected

and voltage controlled phase shifters are used, this step is performed automatically by the

instrument and can be omitted.

3. Finally, the actual measurement can be started by pressing the green “Measure” button as shown

in Figure 18. Measurement can be stopped and restarted without need of performing step 1 and 2

each time.

Figure 17 Additive phase noise measurement tab (Time Domain) during step 2, automatically

adjusting the phase to quadrature.