5

CMAG 012014

Commis-
sioning

Stirring

Heating

Contact-
thermometer
connection

MS 4

MS 7

MS 10

HS 4

HS 7

HS 10

HP 4

HP 7

HP 10

Put device switch (A) in the OFF position

Plug in (G) mains cable

Once connected to the power supply the device is in “stand-by” mode

The right decimal point on the display (C) is lit

Commissioning

Put device switch (A) in the ON position

Any set values are retained when device is switched off and even after the device is

disconnected from the power!

Set the engine speed with the operating button (E) on the right

Put device switch (A) in the ON position

Any set values are retained when device is switched off and even after the device is disconnected from the power!

Set the target temperature for the heating plate using operating button (D) [for VHP (E)]

The set value is indicated on the display (C). If energy is being supplied to the heating plate,

the red LED (B) lights up

In the stir and stand-by modes, “hot” flashes on the display after the heat has been switched off

for as long as the temperature of the set-up surface exceeds 50°C

Put device switch (A) in the OFF position, unplug contact plug

Safety contact thermometer acc. to DIN 12878 class 2 connected with jack (H)

Put device switch (A) in the ON position

Beware the instruction manual of the contact thermometer

Important: Display (C) also indicates the target temperature of the heating plate when

the contact thermometer is connected