Display

In the “Display“ menu the user can specify what information will be
displayed on the main screen.

Note: If the ”Torque”option is activated, by pressing the ”Back“ but-
ton the user can reset the current torque to 0 Ncm as a reference value.
At the same time the Δ appears in front of the Ncm unit.

Programs

Under menu "Programs", 5 user-defined speed (rpm)-time profiles
can be created. In addition, the user could define the intermittent
mode is activated or not in the programs.
If the intermittent mode is activated, the run time/stop time value
from "Intermittent Mode" setting will be taken.
Note: If user need to activate the intermittent mode in one segment
of a program, he should set the "Run time/Stop time" in menu op-
tion "Interval", meanwhile activate the "Run/Stop" function in menu
option "Intermittent Mode" (see section "Stirring").
Once a programs has been selected, the following options are available.

Start: Start the selected program upon request.
When the selected program is started by pressing on menu option
"Start" with rotary/push knob, screen reminds the user to confirm
the program control. Press "OK" with rotary/push knob to start the
program control.

Edit: Edit the selected program parameters.
Start to edit the selected program parameters by pressing on menu
option "Edit" with rotary/push knob. The user can edit, delete or in-
sert one selected program segment in the program.
When user edits the program time for at least one segment, tick (

)

for respective program will appear.

Delete: Deletes the selected program.
If a selected program is deleted by pressing on menu option "Delete"
with rotary/push knob, all the program parameters will be emptied. The
tick (

) disappears.

Operating Mode

Operating Mode A:
In this operating mode, the set speed is not saved when the current
run comes to an end or the device is switched off.

Operating Mode B:
In this operating mode, the set speed is saved when the current run
comes to an end or the device is switched off, and the value can be
changed.

Operating mode C:
In this operating mode, the set speed is saved when the current run
comes to an end or the device is switched off, and the value cannot
be changed.

Details for editing the program:
When a program is being edited, following screen appears.

PROGRAM 1

Seg
No

.

rpm

Time
hh:mm

Interm.
Mode

Edit Delete Insert

1

2

3

4

5

0

00:00

No

In this program, user can define up to 5 seg-
ments. The selected segment is highlighted.
Then, the user can edit, delete or insert a seg-
ment in this program.
When the "Back" button is pressed after edit-
ing, the program is save automatically.

Edit:
When the highlighted "edit" symbol appear on the right top of the
screen, the user could change the speed (rpm), time value or inter-
mittent mode setting.

Note: The speed (rpm) value can be changed during the speed limit
range of the device. The time can be set from 1 minute to 10 hours.

Delete:
When delete a highlighted segment, the settings will be emptied. The
section jump to next segment.

Insert:
With the insert option, a new segment will be insert below the selected
segment. The default settings for the new segment are as following:
rpm: 0
Time (hh:mm): 00:01
Intermittent Mode: No

View the program in graph:
After a program is being edited, press the "Back" button to save the set-
tings. Press the "Back" button again, the screen back to follow screen.

Programs

Start

Edit

Delete

Program 1

Program 2

Program 3

Program 4

Program 5

If no key or knob is pressed or turned for 5 seconds, pictorially respective
program will be shown.

rpm

500

1000

Seg No.:

1

rpm: 150
Time (hh:mm): 00:10
Interm. Mode:

Yes

PROGRAM 1

Detail

rpm

500

1000

PROGRAM 1

Detail

rpm

500

1000

Seg No.:

2

rpm: 150
Time (hh:mm): 00:10
Interm. Mode:

No

PROGRAM 1

Detail

Press the rotary/push knob on option "Detail", the detailed settings

of every segment can be seen.

Turn the rotary/push knob, the details screen of every segment can