Maintenance and cleaning

The instrument is maintenance-free. It is only subject to the natural
wear and tear of components and their statistical failure rate.

Cleaning

For cleaning disconnect the main plug!

Use only cleaning agents which have been approved by IKA

®

to

clean IKA

®

instruments.

Dirt

Cleaning agent

Dyes

isopropyl alcohol

Construction materials

water containing tenside/isopropyl alcohol

Cosmetics

water containing tenside/isopropyl alcohol

Foodstuffs

water containing tenside

Fuels

water containing tenside

For materials which are not listed, please request information from
IKA

®

application support.

Wear protective gloves during cleaning the instruments.

Electrical instruments may not be placed in the cleansing

agent for the

purpose of cleaning.

Do not allow moisture to get into the instrument when cleaning.

Before using another than the recommended method for cleaning or
decontamination, the user must ascertain with IKA

®

that this meth-

od does not destroy the instrument.

Spare parts order
When ordering spare parts, please give:
- machine type
- serial number, see type plate
- item and designation of the spare part see www.ika.com,

spare parts diagram and spare parts list

- Software version.

Repair
Please send in instrument for repair only after it has been
cleaned and is free from any materials which may constitute
a health hazard.
For repair, please request the “Decontamination Certificate” form
IKA

®

or use the download printout of it from IKA

®

website: www.

ika.com.
If you require servicing, return the instrument in its original pack-
aging. Storage packaging is not sufficient. Please also use suitable
transport packaging.

Error codes

The fault is shown by an error message in the display (C) as following if the error occurs, e.g. Error 4.

Proceed as follows in such cases:

Switch the device switch (A) off.

Remove the stirrer tool and remove the instrument from the assembly.

Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.

Error

Cause

Effect

Solution

Error 2

Motor current sensor fault

motor off

- Switch off the instrument

Error 3

Temperature inside instrument is too high

motor off

- Switch off the instrument and allow it to cool down

Error 4

Motor blockage or overload

motor off

- Switch off the instrument
- Decrease the load on the motor and restart again

Error 8

Speed sensor fault or overload

motor off

- Switch off the instrument

Error 21

Safety relay fault

motor off

- Switch off the instrument

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.