4 installation, Installation – Eppendorf PatchMan NP 2 User Manual

Page 24

Advertising
background image

69

Note: The connections between the control board, the mains power supply unit and the motor
module must only be plugged and unplugged with the device switched off.

Note: Do not place the control board down resting on the joystick! When connecting cables,
place the control board on its side on an even surface and secure it against toppling over!

–

The connection to the motor module is made via the corresponding connectors, as shown
in Figure 4.

–

Check the voltage and frequency against the rating plate before connecting the mains
power!

–

Use only the mains power supply unit supplied by Eppendorf!

–

The cable is plugged into the mains power supply unit before being connected to the mains.

Important:

The main power connection is the last one to be connected.

–

Check that the connectors are firmly attached.

Fig. 4: Underside of control board

1 Connection for optional foot switch (position change) (SER2; SubD9, male)
2 Connection (SER1; SubD9, male) (not used on the PatchMan

NP 2)

3 Mains power supply connection

(Important:

Connect only when all the other components have been connected.)

4 Connection for motor module (SubD25, female)
5 Serial port (SER0; SubD9, male) for computer (RS 232)

4.4 Connecting cables – Startup

1

2

3

4

5

4 Installation

Installation

4

04_Install_e.fm Seite 69 Dienstag, 16. August 2005 3:17 15