5.2.4 Capillary exchange
Before swivelling it out, make sure the capillary cannot be impeded by any obstacles!
Wear goggles to protect against possible glass splinters! Swivel out module unit on
Y/Z module!
Swivel module unit for capillary exchange out of working radius. The capillary swivels out
forwards and upwards with the module unit.
–
Before changing the capillary, bring the capillary to a position over the petri dish either
manually or by way of the Home function.
–
Grip module unit at rear motor unit (Y/Z module) and swivel out.
–
Do not use the X module!
–
Exchange capillaries and swivel module unit back into place.
–
Bring the new capillary back into the focal plane.
Fig. 5: Module unit in working position and swivelled out
(here without capillary holder/preamplifier)
0
40°
40°
20°
20°