System management
Mastercycler pro
English (EN)
59
Performing a
software
synchronization
Use the
update
node to perform a software synchronization between the cycler and the
control panel. The software synchronization may be necessary if you receive error
messages after updating the software of the control panel.
1. Highlight
Update
node.
2. Press
Cycler
function key.
3. Select the Cycler.
4. Press
Start
.
Exporting
user-related data
Use the
UserLog
node to export or display all the actions that were performed on the
control panel as a log file.
You can save login and logout information, all changes and interventions in a running
program related to specific users.
Mark
UserLog
and press the desired function key:
Show
: Display user log. You can print or export the user log in the display.
Exp+Clear
: Transfer the user log to the MMC or USB storage medium and delete
the log from the control panel afterwards.
Export
: Copy the user log to the MMC or USB storage medium.