7 transferring programs from the mastercycler 533x, Transferring programs from the mastercycler 533x – Eppendorf Mastercycler pro User Manual
Page 64

System management
Mastercycler pro
English (EN)
64
3. Select
Software Update Control Panel
entry and press the
Start
function key.
The software version currently installed and the software version on the storage medium are displayed.
4. Press the
Update
function key to start the update. You can cancel the operation using the
Restart
function key.
The update may take several minutes. Afterwards, the control panel will automatically restart.
8.7
Transferring programs from the Mastercycler 533x
PCR programs established on the Mastercycler (5333) or the Mastercycler gradient (5331) can easily be
transferred to the Mastercycler pro without first having to optimize the temperatures and hold times.
Enter the program into the Mastercycler pro editor:
Example
1 T = 94.0°
0:02:00
2 T = 94.0°
0:00:15
3 T = 60.0°
0:00:15
+0.0° +0:00
R = 3.0°/s
+0:0°/s
G =10.0°
4 T = 72.0°
0:00:30
5 GOTO 2
REP 29
6 T = 72.0°
0:02:00
7 HOLD 4.0°
ENTER