7 transferring programs from the mastercycler 533x, Transferring programs from the mastercycler 533x – Eppendorf Mastercycler pro User Manual

Page 64

Advertising
background image

System management
Mastercycler pro
English (EN)

64

3. Select

Software Update Control Panel

entry and press the

Start

function key.

The software version currently installed and the software version on the storage medium are displayed.

4. Press the

Update

function key to start the update. You can cancel the operation using the

Restart

function key.

The update may take several minutes. Afterwards, the control panel will automatically restart.

8.7

Transferring programs from the Mastercycler 533x

PCR programs established on the Mastercycler (5333) or the Mastercycler gradient (5331) can easily be
transferred to the Mastercycler pro without first having to optimize the temperatures and hold times.

 Enter the program into the Mastercycler pro editor:

Example

1 T = 94.0°

0:02:00

2 T = 94.0°

0:00:15

3 T = 60.0°

0:00:15

+0.0° +0:00

R = 3.0°/s

+0:0°/s

G =10.0°

4 T = 72.0°

0:00:30

5 GOTO 2

REP 29

6 T = 72.0°

0:02:00

7 HOLD 4.0°

ENTER